About the knowledge of reliability test equipment
although the reliability test equipment () industry is a traditional industry, and it is not too late to start in China's science, technology, national defense and other industries, the real pendulum pre lift angle: 150 ° is still popular in the past 20 years or so. Although there are currently no less than 300 production and sales companies of all sizes in China, few companies really understand and master the core and essence of products. For users, the product looks like a seller's market, and there are many choices, but because there are not many highly skilled users who understand the use performance of the product, coupled with the so-called prepayment system of the industry's transportation and marketing operation, the so-called "sale pending production mode" and other tricky, as well as the dealers playing with the text games on technical indicators, the marketing history of environmental test products in the domestic market for many years is the history of users' blood and tears. Hehe, it's over! I'm engaged in technical work, and I don't want to say more about other things, but what I can't see is that many sample technical schemes use index configuration to engage in covestro. Covestro is one of the world's largest producers of polymeric materials, and the distributors deceive users. The most pitiful thing is that new users have no way to learn. It's hard to buy expensive or poor, and they continue to grow in the middle of the market
therefore, I will explain some indicators of the sample scheme with my understanding, hoping to be useful to the majority of users. Of course, there are many environmental test index projects, and the configuration and operation methods also have their own advantages. If we want to talk more comprehensively, we can publish textbooks. Our personal ability is limited, and we dare not take risks. I can only make a shallow explanation of the commonly used temperature and humidity indicators and the currently commonly used configurations in my cognitive range to help users who are not deep in Taoism get started
performance index analysis
(I) temperature index
1) temperature range: refers to the limit temperature that the product studio can withstand and/or reach. It usually contains the concept of constant control, which should be the extreme value that can operate stably for a relatively long time. The general temperature range includes extreme high temperature and extreme low temperature. The temperature range of the pure high-temperature box is relatively simple. The low temperature is generally at room temperature plus about 10~50 ℃, and the high temperature is at nominal temperatures such as 200, 300, 400 ℃
extreme high temperature of high and low temperature box, nominal temperature of European and American products is about +180 ℃, estimated from the sealing material; The nominal temperature of Japanese and Taiwan products is +150 ℃, which should be a reliable index that the temperature field index (deviation, etc.) can generally meet; Yinhe Sida and other samples have a nominal temperature of +100 ℃, and the technical agreement can actually be signed at about +130 ℃, from thermal insulation materials. Of course, according to user requirements, most of the company's technical agreement indicators can be signed to +180 ℃, but whether the limit temperature test can be carried out has to be verified in kind. I suggest that if it is not necessary, the high temperature of 130 ℃ is more appropriate. Extreme low temperature is mainly determined by refrigeration capacity and refrigeration mode. Here, only the indicators of mechanical compression refrigeration system configuration are introduced. Generally, the limit capacity of unipolar refrigeration is -45 ℃, and the general index is set at -40 ℃; Theoretically, the limit capacity of cascade refrigeration can reach -82 ℃, but considering that the refrigerating capacity below -70 ℃ is too small, the general index is set at -70 ℃. Since some users do not need to use lower temperatures, most companies will launch -1 with higher configuration cost performance The photoelectric encoder transmits the measured speed signal to the spindle speed measurement control system and the spindle speed measurement display system for 20 ℃ or even 0 ℃. As for the Taiwan funded sample index of every 10 ℃ level, I think it is a desktop commercial deception, of course, there are differences in technology
2) temperature fluctuation: this index is also called temperature stability. After controlling the temperature to be stable, the difference between the highest and lowest temperature at any point in the workspace within a given time interval. There is a small difference here. The working space is not a studio, but a space about 1/10 of the length of each side of the studio from the box wall. This indicator assesses the control technology of the product. The index required by the general standard is 1 ℃ or 0.5 ℃. In order to boast that the control technology has a nominal temperature of 0.2 ℃, in fact, the temperature fluctuation of temperature test control is less than 0.3 ℃, and the temperature uniformity and temperature deviation below are difficult to pass; It will be difficult to control the temperature fluctuation in the humidity test. If the temperature fluctuation is less than 0.1 ℃, it will be difficult to stabilize the humidity. It should be noted that there are temperature stability indicators such as air aging test chamber, but it is not a concept
3) temperature uniformity: the old standard is called uniformity, and the new standard is called gradient. After the temperature is stable, the maximum value of the difference between the average values of temperature at any two points in the workspace at any time interval. This index can assess the core technology of products better than the following temperature deviation index, so many company samples and plans deliberately hide this item. The general standard requires that the index is 2 ℃, but objectively speaking, it is uneconomical for users to maintain the uniformity of pure high-temperature test chamber above 200 ℃
4) temperature deviation: the difference between the average temperature in the center of the workspace and the average temperature at other points in the workspace at any time interval after the temperature is stable. Although the old and new standards have the same definition and title for this indicator, the detection has changed. The new standard is more practical and more rigorous, but the assessment time is shorter
the general standard requires that the index is 2 ℃, and the actual operating temperature (℃) of pure high-temperature test chamber above 200 ℃ can be 2%. There is no temperature uniformity and temperature deviation called by the mainland in the indicators of aisli Taiwan capital. It is similar to the uniformity of temperature distribution. The English words are also different. The sample indicators are 0.5 ℃, 1 ℃, 1.5 ℃, etc. Hope to explain to friends who know about the logo
5) temperature change rate (): This is an indicator to assess the product configuration capacity. The forms given by various companies are also diverse, including temperature rise and fall speed, temperature rise and fall time, heating and cooling capacity, etc., and the temperature rise and fall range is not unified. Of course, there is also the most fishy. Generally, users who don't require temperature changes mainly specify the time from normal temperature to extreme low temperature. Although it's as big as Guangai and a little tricky (strictly speaking, it's professional), it doesn't matter if it's 3-5 minutes away. For the test with temperature change requirements, when ordering the test chamber, the easiest way for the user is to repeat this several times until cleaning in the signing of the technical agreement. It is required to first clarify the load test, and then list the volume, size, weight, basic material composition, possible maximum calorific value and other indicators of the test sample, as well as the starting and ending temperature points of the relevant test standards The temperature change curve of the process description is all written, so that the supplier has no loopholes to drill. Don't take the sample index as the agreement index, because the interpretation right of the sample index lies with the supplier. The test according to the newly revised national standard temperature change rate is no-load, and the second is that you can give a 20% discount when calculating (thinking of Hou Baolin's crosstalk selling cloth). Even the sample index such as Fuchi ACS is very difficult for non professionals to read. Aisli Testing Instrument Co., Ltd., consulting:, address: